Animal Anesthesia ventilator matters needing attention
1 anesthesia ventilator when using oxygen pressure in anesthesia ventilator pressure meter green scale range, the general use of 0.4Mpa.
Anesthetist is generally used to looking at oxygen decompression tables pressure regulation, but some oxygen decompression table due to crash, the use of long, leading to pressure readings are not allowed, suggested that the oxygen pressure regulation should also pay attention to anesthesia ventilator comes with pressure gauge is normal. Oxygen pressure and normal use of anesthesia respirator is very important, if the pressure is too high, easily washed away the components inside the machine, resulting in pipeline off; if the pressure is too low, especially for the pneumatic control anesthesia respirator, because it uses gas as power. If the pressure is too low, will lead to machine moisture output is low or even no tidal volume output.
2 need to unplug the oxygen decompression tables, avoid using brute force, the outer ring should be pressed under the oxygen outlet, then pulled out. 3 before use to check whether the flow control flow control button in full scale free; rapid switch can in 7-10s filled with oxygen
Man's breathing bag; air tightness is good.
4 flow sensor if the water will affect the accurate measurement of tidal volume. Flow sensor is generally used to test the platinum gold wire
Volume, if the water should be soaked in 75% alcohol disinfection after take out to dry naturally, not with strong winds dried and dried with a cloth, otherwise it may caused by platinum wire fracture.
5 anesthesia ventilator used to note the power connection is good, observe the power supply indicator is AC power or
Battery power supply, so as not to run out of battery power consumption and even damage caused by the inconvenience.
Anesthetic drugs 6 evaporation tank can only fill the same; not allowed in an evaporation tank using two or more kinds of narcotic drugs, not allowed
On the same time, the two evaporation can be used simultaneously on the anesthesia ventilator with the interlocking device, and the evaporation tank is not allowed to be removed in the use of the device.
2016.4.26