A maintenance cycle
1. Every day to clean the external surface.
21% oxygen concentration calibration with 2 weekly (oxygen sensor breathing circuit). 3 every two weeks emptying
and discard the anesthetic vaporizer.
100% oxygen concentration calibration with 4 per month (the oxygen sensor of the breathing circuit). 5 every year
to replace the external O rubber ring evaporation tank on the port. 6 on-demand:
The installation of the new cylinder gasket cylinder yoke. The emptying water collector, replace the absorption in
the tank.
The replacement of the oxygen sensor breathing circuit (in normal use, the sensor performance in accordance with
the specifications in a year).
Replacement of the disposable plastic flow sensor (under normal circumstances, the performance of the flow sensor
at least within six months in line with technical specifications).
2.Daily maintenance.
Before 1. Daily use to check for breathing circuit connection is correct, intact and respiratory system containing
sufficient absorbent; and detection of evaporation tank installation. 2 breathing loop system to be kept in a clean
and dry environment, away from light and heat. Avoid contact with metals, organic solvents, oils or fats and
strong cleaning agents.
3 breathing circuit cleaning: first removal of organic substances and chemical residues in products or disinfected
before autoclave sterilization. Steam high pressure conditions, the maximum recommended temperature of 134
degrees Celsius, the time is not more than 20mins. Can be used before each patient or according to the actual use
of the situation, the need to increase the number of disinfection.