The infusion pump fault discharge method

The door tight of infusion pump:
In between the upper and lower shaft with lubricating oil, then repeatedly opened and closed until the switch freely, if not rule out the possibility of failure, then opens the door assembly, with fine sandpaper grinding shaft and cleaning shaft hole after reinstall door assembly.  
Note:
1, installed after the door must install stopper.
2, pay attention to the spring plate installation direction.  
3, when changing the probe cable to pay attention to do not damage the insulation of the skin, it is certain that the plug is shielded wire grounding.
4, the installation bubble emission probe don't let the upper platen extrusion cable.  
5, check and adjust the bubble gap.  

6, check the pump pressure and accuracy.